Digital Glass Body pH Electrode for Hydrocarbons and Solvents - HI10430

£

Digital Glass Body pH Electrode for Hydrocarbons and Solvents - HI10430

 

Details

digital electrode

Built-In Microchip

 

The built-in microchip stores sensor type, serial number and calibration information including date, time, offset, slope, probe condition and buffers used. This information is automatically retrieved by edge® once the electrode is plugged in. The ability to transfer information allows for hot swapping of probes without having to recalibrate. 

All pH measurements are performed within the electrode and transferred digitally to the meter. This overcomes any noise issues associated with the traditional high impedance analog measuring system. Electrical noise can be generated from a built-in temperature sensor and while working in a humid environment.

 

spherical-tip

Spherical Glass Tip

The spherical shaped tip design allows for a wide area of contact with the measured sample. This permits a faster electrode response with a higher degree of stability. The glass sensing tip on the HI10430 uses Hanna’s high temperature (HT) glass formulation that has a resistance of approximately 400 megaohms. The high impedance makes the HI10430 ideal for aggressive chemical applications and high temperature applications. Conventional glass formulations suffer from a greatly reduced lifespans at high temperatures relative to high temperature glass.

 

spherical-tip

Triple Ceramic Junction

The triple ceramic junction allows a higher flow rate of electrolyte from the reference cell into the solution. A standard pH electrode will use a single ceramic junction that allows for 15 to 20 µL/hour of electrolyte to flow; the HI10430 has three ceramic junctions, providing for 40 to 50 µL/hour of electrolyte to flow. This high flow rate provides faster electrode response and a more stable measurement in viscous solutions or samples of low conductivity.

 

open junction

Double Junction Reference

A double junction electrode has an internal compartment surrounding the reference wire. Silver ions are present in the electrolyte of the internal compartment, which houses the Ag/AgCl reference wire; the electrolyte outside this compartment is silver free. The double junction design means that virtually no silver from the electrode enters the sample. This design allows measurement in applications where silver ions in the sample are undesirable or silver precipitates on the junction are likely to form.

 

glass-body

Glass Electrode Body

The glass electrode body is suitable for a wide range of applications due to its chemical resistance. The glass electrode is compatible with many non-aqueous solvents and other aggressive chemicals. Glass is also resistant to many forms of radiation, such as ultraviolet radiation.

 

 

Measurement Range pH: 0 to 13
Reference Cell Type double
Junction Type/Flow Rate ceramic, triple / 40-50 μL/h
Refill Electrolyte 3.5M KCl
Body Material glass
Tip / Shape spheric (dia: 9.5 mm)
Matching Pin no
Max. Pressure 0.1 bar
Temperature Operating Range -5 to 100°C (23 to 212°F)
Body Length/Overall Length 120 mm / 175.5 mm
Temperature Sensor yes
Outer Diameter 12 mm
Cable coaxial; 1 m (3.3’)
Connector Type 3.5 mm connector
Amplifier no
Recommended Use hydrocarbon, low conductivity solutions, paints, sea water, solvents, strong acid and base, tris buffer

 