AD-4212C Series Weigh Modules

£

AD-4212C Series Weigh Modules

External Input/Output
RS-232C interface
Accessories
AD-1683 : Static eliminator
AD-1684A : Electrostatic fieldmeter
AD-1687 : Weighing environment logger
AD-1689 : Tweezers for calibration weight
AD-8118C Universal printer
AD-8127 : Compact printer
AD-8526 : Serial/Ethernet converter
AD-8922A : Remote display unit
AD-8923-BCD : Remote controller (BCD)
AD-8923-CC : Remote controller (CC-Link)
AX-USB-9P : Serial/USB converter with cable
Specifications
Model AD-4212C-301* AD-4212C-3100*
Capacity Precision range 51 g 510 g
Standard range 320 g 3200 g
Readability Precision range 0.0001 g 0.001 g
Standard range 0.001 g 0.01 g
Stabilization time (when set to
FAST under a good environment)
Precision range 1.3 sec.
Standard range 1.0 sec.
Repeatability (Standard deviation) Precision range 0.0002 g 0.002 g
Standard range 0.001 g 0.01 g
Linearity ±0.002 g ±0.02 g
Sensitivity drift (10 °C to 30 °C) ±2 ppm/ °C
Weighing pan 50 mm × 50 mm
* Smart range function: The display will switch to the standard range automatically when the value exceeds 51 g / 510 g but returns to the precision range by performing RE-ZERO (tare).
Model AD-4212C-300 AD-4212C-600 AD-4212C-3000 AD-4212C-6000
Capacity 320 g 620 g 3200 g 6200 g
Readability 0.001 g 0.01 g
Stabilization time (when set to
FAST under a good environment)
0 - 30 g 0.5 sec. 0 - 30 g 0.5 sec. 0 - 300 g 0.5 sec. 0 - 300 g 0.5 sec.
30 - 320 g 1.0 sec. 30 - 620 g 1.0 sec. 30 - 3200 g 1.0 sec. 30 - 6200 g 1.0 sec.
Repeatability (Standard deviation) 0.001 g 0.01 g
Linearity ±0.002 g ±0.005 g ±0.02 g ±0.04 g
Sensitivity drift (10 °C to 30 °C) ±2 ppm/ °C
Weighing pan 50 mm × 50 mm
CAD AD-4212C CAD File (DXF)
Just 5 mg after 30 million times!
 AD-4212C-300 (320 g × 1 mg) completed an endurance test
 of 30 million repeated loadings, which took more than 1 year.
 The final value of span drift was a mere 5 digits, or 5 mg (display value
 of 200.005 g for a standard weight of 200 g), demonstrating that the
 product can maintain its accuracy over a long period of use.
 See how the test was performed here.
Development Story
Dedicated software WinCT-AD4212C (Ver.1.13)
PDFAddition to Manual for WinCT-AD4212C (PDF 245KB) 
This manual is for users who wish to adjust the response settings of the AD-4212C Electronic Balance beyond the standard settings.
PDFBrochure (PDF 1.45MB)
PDFManual (PDF 1.41MB) 
PDFAD4212C material table (PDF 5KB)
Demonstration of fast stabilization (YouTube)