Manganese High Range Portable Photometer - HI96709

£

Manganese High Range Portable Photometer - HI96709

Details

The HI96709 portable photometer is for the high range measurement of manganese. Hanna's portable photometers feature an advanced optical system; the combination of a special tungsten lamp, a narrow band interference filter, and silicon photodetector ensure accurate photometric readings every time. The Hanna exclusive CAL Check feature utilizes ready-made, NIST traceable standards to verify both meter validation and calibration. The exclusive cuvette locking system ensures that the cuvette is inserted into the measurement cell in the same position every time to maintain a consistent path length.

 

Features at-a-glance

CAL Check - Allows for performance verification and calibration of the meter using a CAL Check NIST traceable secondary standard.

GLP - Records the date of the last user calibration performed.

Built in timer - Allows the use of a timer to ensure the appropriate 3 minutes and 30 seconds reaction time is used for the chemical reaction. At the end of the timer countdown the meter will automatically take the reading. This feature ensures consistency across multiple users.

Indexing indent on meter - The meter has an indent that allows for a lock and key fit with the cuvette cap. This ensures that the cuvette is indexed consistently (same position) in order to maintain the same path length for accurate results.

Cooling Lamp Indicator - It is necessary to maintain a consistent temperature of the optical components in order to maintain a narrow wavelength band of light. This photometer has a cooling lamp indicator that is displayed for a short period of time before each measurement in order to allow the components to cool and obtain the highest accuracy possible.

Error messages - Messages on display alerting to problems including no cap, over range and under range readings, and light source error.

Auto-shut off - Automatic shut off after 10 minutes of non-use when the meter is in measurement mode. Prevents wastage of batteries in the event the meter is accidentally not turned off.

Battery status indicator

The instrument has a battery level indicator to display reaming life as follows:

  • 3 lines for 100 % capacity
  • 2 lines for 66 % capacity
  • 1 line for 33 % capacity
  • Battery icon blinks when the capacity is less than 10 %.

Units of measure - Appropriate unit of measure is displayed along with reading.

 

Manganese is one of the most common metals present in nature and is used in many industrial applications, for example, the production of fertilizers and in the pharmaceutical industry. Manganese salts are also used in iron alloys (steel manufacturing) and non-iron alloys as it improves their corrosion resistance and hardness.

The HI96709 uses an adaptation of Standard Methods for the Examination of Water and Wastewater, 18th edition, periodate method to measure manganese concentrations of less than 20.0 mg/L (ppm) manganese. When the reagent is added to samples containing manganese, the sample will turn a violet hue; the greater the concentration, the deeper the color. The associated color change is then colorimetrically analyzed according to the Beer-Lambert Law. This principle states that light is absorbed by a complementary color, and the emitted radiation is dependent upon concentration. For manganese determination, a narrow band interference filter at 525 nm (green) allows only green light to be detected by the silicon photodetector and omits all other visible light emitted from the tungsten lamp. As the change in color of the reacted sample increases, absorbance of the specific wavelength of light also increases, while transmittance decreases.

 

The HI96709 is available as a kit (HI96709C) which includes:

  • Portable Photometer
  • Sample cuvettes (2)
  • Caps
  • CAL Check™ standards with certificate
  • Cuvette cleaning cloth
  • Scissors
  • Rugged carrying case
    • Carrying case has thermoformed insert for meter and accessories

*Reagents ordered separately

CAL Check™ standards with certificate

The HI96709-11 CAL Check standards are used for calibration and performance verification of photometers with the CAL Check function.

Supplied with Certificate of Analysis

  • Lot number
  • Expiration date
  • Standard value @ 25 °C
  • Reference meter NIST traceable

Provided storage containers

  • Light tight
  • Protects from accidental breakage
Manganese Range 0.0 to 20.0 mg/L (ppm)
Manganese Resolution 0.1 mg/L (ppm)
Manganese Accuracy ±0.2 mg/L ±3% of reading
Manganese Method adaptation of Standard Methods for the Examination of Water and Wastewater, 18th edition, Periodate method
Photometer/Colorimeter Light Source tungsten lamp
Photometer/Colorimeter Light Detector silicon photocell with narrow band interference filter @ 525 nm
Battery Type/Life 9V battery
Automatic Shut-Off after ten minutes of non-use in measurement mode; after one hour of non-use in calibration mode; with last reading reminder
Environment 0 to 50°C (32 to 122°F); RH max 95% non-condensing
Weight 360g (12.7oz.)
Dimensions 193 x 104 x 69 mm (7.6 x 4.1 x 2.7”)
Ordering Information HI96709 is supplied with sample cuvettes (2) with caps, 9V battery, instrument quality certificate and instruction manual. CAL Check™ standards and testing reagents sold separately.