Conductivity Mini Controller (0 - 1999 μS/cm) - BL983313

£

Conductivity Mini Controller (0 - 1999 μS/cm) - BL983313

 

Details

BL983313

Adjustable Dry Contact Dosing Relay

The BL983313 features a dosing relay that is activated when the reading is above a user programmable set point.

Programmable Overdose Protection

For enhanced safety the mini controller can be programmed to deactivate the dosing relay if the set point is not reached within a specified time interval. The overdosing timer is programmable from 5 to 30 minutes or disabled.

Relay Control Override

With the flick of a switch the mini controller's relay can be disabled (Off), placed in control (Auto), or be activated for manual operation (On) which is useful for priming a dosing pump.

Multicolor LED Indicator

Multicolor LED indicator allows an operator to quickly check the status of the controller.

  • Green = Meter in measurement mode and reading is below the set point.
  • Orange/Yellow = Reading is above the set point and the relay is activated.
  • Blinking Red = Indicates an alarm condition such as when the maximum dosing time has been exceeded.
BL983313

Fuse Protected Dosing Contacts

The relay dosing contact is rated for up to a 2A load and is fuse protected.

Labeled Termination Connections

Quick-connect terminal blocks are clearly labeled for easy connection to power, conductivity probe and relay that can be used to operate a dosing pump, valve, audible alarm, or light.

HI7634

1/2" NPT Male Threaded Amperometric Probe for Mounting In-line*

The BL983313 mini controller uses the HI7634 conductivity probe that is available with different cable lengths.

*HI7634 is not included and must be ordered separately

 

BL931700

Front View

Installation cut-out dimensions are provided as well as outside panel dimensions.

BL931700

Side View

Adjustable location brackets allow the controller to slide easily and securely into place..

Range 0 to 1999 μS/cm
Resolution 1 μS/cm
Accuracy (@25°C/77°F) ±2% f.s.
Calibration manual, with CAL trimmer
Dosing Relay maximum 2A (fuse protected), 250 Vac, 30 VDC
Dosing Selection contact closed when measure > setpoint
Setpoint adjustable from 0 to 1999 μS/cm
Overtime adjustable, typically from 5 to approximately 30 minutes
Temperature Compensation automatic from 5 to 50°C (41 to 122°F) with β =2%/oC
Power Supply models “-0”: 12 VDC adapter (included)
models “-1”: 115/230 VAC; 50/60Hz
Dimensions 83 x 53 x 99 mm (3.3 x 2.1 x 3.9’’)
Weight models “-0”: 200 g (7.1 oz.) models “-1”: 300 g (10.6 oz.)
Ordering Information BL983313-1 (115/230 VAC) and BL983313-0 (12 VDC) are supplied with mounting brackets, transparent cover and instruction manual.

 