HI7091L Pretreatment Reducing Solution for ORP electrodes (500 mL)

£

HI7091L Pretreatment Reducing Solution for ORP electrodes (500 mL)

Details

ORP (Oxidation Reduction Potential) electrodes are used to measure a chemical reaction in which one species is being oxidized and another is being reduced. The loss of electrons by an element or compound is known as oxidation reaction while the gaining of electrons by an element or compound is known as reduction reaction. For example, plating companies that have hexavalent chromium (Cr6+) in their wastewater must reduce it to a safer trivalent chromium (Cr3+) form. This reduction reaction occurs in the presence of a strong reductant such as sodium metabisulfite. The addition of the reductant must be measured with an ORP electrode to ensure that there is an adequate reducing potential. Using the HI7091L reducing pretreatment solution prior to taking a measurement provides for a faster response time in a reducing environment.

To use, soak the ORP electrode in the pretreatment solution for 15 minutes prior to use. As a general rule, if the mV reading corresponding to the pH measurement is lower than the value in the table below the reducing pretreatment solution (HI7091) is used. If the mV reading corresponding to the pH measurement is higher than the value in the table below then the oxidizing pretreatment (HI7092) should be used.

    

pH Value

mV Value

pH Value

mV Value

pH Value

mV Value

pH Value

mV Value

pH Value

mV Value

0

990

3

800

6

640

9

460

12

280

1

920

4

740

7

580

10

400

13

220

2

860

5

680

8

520

11

340

14

160



The HI7091 bottle is labeled with

  • Product code
  • Lot number
  • Expiration date

 

Air tight bottle with tamper-proof seal

  • Ensures quality and freshness of solution
Description reducing pretreatment solution
Certificate of Analysis no
Package bottle
Quantity 1
Size 500 mL