Digital Refractometer for Sugar (% Brix) Analysis in Wine, Must and Juice - HI96811

£

Digital Refractometer for Sugar (% Brix) Analysis in Wine, Must and Juice - HI96811

 

Details

The HI96811 Digital Wine Refractometer converts the refractive index of a wine, juice or must sample to % Brix. This conversion is based on the tables found in the ICUMSA Methods Book (International Commission for Uniform Methods of Sugar Analysis) that documents the changes in refractive index with temperature for a percent by weight sucrose solution. Since the majority of sugar in grape juice and must is fructose and glucose instead of sucrose, the reading is sometimes referred to as “Apparent Brix”. Typical grapes at harvest will be between 19-24% Brix or degrees Brix (°Bx). The HI96811 is an easy-to-use tool for winemakers for measuring sugar in grapes or must, in the field or in the lab.

  • Designed for wine sugar analysis
  • Temperature Compensation algorithms based on sucrose solution
  • 0 to 50% Brix range with an accuracy of +/- 0.2% Brix

Making a Standard % Brix (Sucrose) Solution

To make a Brix Solution, follow the procedure below:

  • Place container (such as a glass vial or dropper bottle that has a cover) on an analytical balance.
  • Tare the balance.
  • Weigh out X grams of high purity sucrose (CAS #: 57-50-1) directly into a beaker
  • Add distilled or deionized water to the beaker so the total weight of the solution is 100g.
  • Note: Solutions above 65% may need to be vigorously stirred or shaken and heated in a water bath to roughly 40°C (104°F). Remove solution when sucrose has dissolved and allow to cool before use.
Example to prepare as 25% Brix solution:
 
% Brix Sucrose (g) Water (g) Total (g)
25 25.000 75.000 100.000
 
refractive index

Refractive Index

The HI96811 takes measurements based on a sample’s refractive index. Refractive index is a measurement of how light behaves as it passes through the sample. Depending on the sample’s composition, light will refract and reflect differently. By measuring this activity with a linear image sensor, the sample’s refractive index can be assessed and used to determine its physical properties such as concentration and density. In addition to the linear image sensor, the HI96811 uses an LED light, prism and lens to make measurement possible.

Variations in temperature will affect the accuracy of refractometry readings, so the use of temperature compensation is highly recommended for reliable results. The HI96811 contains a built-in temperature sensor and is programmed with temperature compensation algorithms in accordance with the ICUMSA Methods Book Standard for a percent by weight sucrose solution.

 

Video


 

Features at-a-glance:


One-point calibration
  • Calibrate with distilled or deionized water

Small sample size
  • Sample size can be as small as 2 metric drops (100 μl)

Automatic Temperature Compensation (ATC)
  • Samples automatically compensated for temperature variations

Quick, accurate results
  • Readings are displayed in approximately 1.5 seconds

Dual-level LCD
  • The dual-level LCD displays % Brix and temperature readings simultaneously



Stainless steel sample well
  • Easy to clean and corrosion-resistant

Battery indicators
  • Battery percent level remaining at startup and low battery indicator

Automatic shut-off
  • To conserve battery life the meter shuts off automatically after three minutes of non-use

IP65 water resistant protection
  • Water resistant ABS plastic casing designed to perform under laboratory and field conditions.
Sugar Content Range 0 to 50% Brix
Sugar Content Resolution 0.1% Brix
Sugar Content Accuracy ±0.2% Brix
Temperature Range 0 to 80°C (32 to 176°F)
Temperature Resolution 0.1°C (0.1°F)
Temperature Accuracy ±0.3 °C (±0.5 °F)
Temperature Compensation automatic between 10 and 40°C (50 to 104°F)
Automatic Shut-Off after three minutes of inactivity
Battery Type/Life 9V / approximately 5000 readings
Dimensions 192 x 102 x 67 mm (7.6 x 4.01 x 2.6”)
Weight 420 g (14.8 oz.)
Refractometer Light Source yellow LED
Minimum Sample Volume 100 μL (to cover prism totally)
Sample Cell stainless steel ring and flint glass prism
Measurement Time approximately 1.5 seconds
Ordering Information HI96811 is supplied with battery and instruction manual
Notes Enclosure Rating: IP65

 